Bridge
Model TD50
Capacitance
Measurement:
Ranges: Four
ranges; Max. scale ranges :
1,1 nF; 11 nF; 110
nF; 1,1 mF with Cs= 100 pF
Scale ranges x10 with
Cs = 1 nF
Resolution: 0,001
pF (at 1 nF range)
Accuracy: ± 0,01
% of scale range
Dissipation
Factor (tg():
Ranges: Three
ranges:
1 % - 10 % - 100 %
Resolution: 0,01 %
of max. scale range
Accuracy: ± (1%
of reading + 0,3 % of scale range) with external
cable length and standard capacity compensation
matched
Maximum current at
1 mF range: 20 A
Sensitivity
Minimum voltage for
guaranteed accuracy:
1000 V with Cs =
100 pF
100 V with Cs = 1 nF
Power frequency:
50 or 60 Hz (must be specified for direct reading)
High voltage
protections included
|
Power
Supply Model FB15
Output
voltage: 0 - 15 kV
Output current:
< 200 mA
UST, GST and GSTg
modes
Meters: digital
kilovolt and milliampere meters
Interference
Suppression Controls: Both module and argument
controls to suppress interference current
Standard Capacitor:
Solid insulation
Capacitance 100pF
- tgd < 0,02 %
Input: 110 / 220 V
- 50 / 60 Hz
Dimensions:
Standard cabinets 19"
TD50: 483 x 208 x
400 mm
FB15: 483 x 430 x 500
mm
Accessory (optional)
Transformer Model
TA-100 to expand range of capacitance up to x
1000
|